8.14. Detector failure (PC 2003)
The detector alarm is activated when:
- The detector high voltage power supply is defective.
- The detector is located in a background radiation field and the low range Geiger is
defective. If the detector is located in a radiation field higher than 800mR/h, and the
meter displays the following readings sequence: 800 mR/h, 600 mR/h, 0.00 mR/h -
this indicates high range Geiger malfunction.
Check (500÷550)V
on R4/R5
Check 5V on C2
Check (500÷550)V
on R4/R5
Check positive
pulses on ZP1201
G1 (Geiger pin)
O.K.
Detector failure
O.K.
O.K.
Yes
O.K.
Replace U3
END
END
Replace Q2 or Q4
Check TTL pulses
on U3/8
Check TTL pulses
on J1/4
Replace U1
Replace T1
Check and fix the
cable between
meter & detector
O.K.
O.K.
O.K.
O.K.
Replace Q3 & Q1
Check TTL pulses
on U3/1
Check (500÷550)V
on R4/R5
Check (500÷550)V
on R4/R5
Replace Card