TelePole WR/ Operating & Maintenance Manual Page 21
6. Setting ID Number
Note: Setting the ID # must be done only by an authorized user.
The ID # includes 6 digits.
The second digit in the ID # sets the instrument type. TelePole’s first digit is (0).
Setting and Checking the ID #:
Unscrew the calibration screw located on the back of the meter. Turn the internal switch to
the calibration position.
The SET LCD's are displayed.
1 - Long press on MODE push-button (5 sec.) - In/out display and set ID #.
2 - Short press on LIGHT push-button - Increases displayed digit value.
3 - Short press on RESET push-button - Displays next digit value.
For example:
The default ID# is:
d1.9
d1.6
0.00
d2.0
Display reading
2
3
1
1
Digit’s value
ID# digit
906325
9 06 001
Id1.9
Id4.3
= ID #
Serial # of Instrument
Year
Designates Instrument type: TelePole
Designates Portable Instrument