TelePole WR/ Operating & Maintenance Manual Page 40
8.13. Instrument does not measure the radiation
Replace or fix cable
Check cable between meter
and detector
System does not measure
the radiation field
No
Yes
O.K.
O.K.
O.K.
O.K.
Replace detector
Check TTL pulses on
U2/5 (PC1945) when
the detector is in the
radiation field
Check TTL pulses
on U8/4
Meter OK?
No
No
END
Yes
Detector failure
See section 8.14
Detector or
cable failure
No
Check 5V on U2/1 (PC1945)
O.K.
Check TTL pulses on TP1
(PC1945)
O.K.
Check and fix wire
connection between
PC1942 & PC1945
Replace CPU card
Replace U2
No
No
No
Check and fix wire
connection between
PC1942 & PC1945