10. Calibration Procedure for the DRM-1/2D
10.1 Preface
T
he DRM-1/2 and the DRM-1D can be calibrated via the RS232 communication port with a NULL modem
adapter.
Switch to the ON position dipswitches 1&2 for RS232 and the calibration will be performed by using the Meter
Calibration software.
If the radiation field is greater than 10 mR/h, the software will automatically enable the setting of the appropriate
factor, depending on the field intensity for the DRM-1/2. Refer to section 10.1 for instructional use of the
calibration software.
10.1.1 Calibration Ranges
F1 in a field > 10 mR/h and < 100 mR/h.
F2 in a field > 100 mR/h and < 800 mR/h.
F3 in a field > 800 mR/h and < 50 R/h.
F4 a field > 50 R/h and < 350 R/h.
F5 in a field > 350 R/h
Note: If the radiation field is lower than 10 mR/h, each factor can be set manually by the
calibration software.
10.2 Operating Instructions
1.
Run the calibration program via the RMC.EXE file (or icon).
2. Enter a Login name/calibrator (‘jsmith’) name that will appear in the calibration documentation
(up to 12 characters).
3. Enter Password – “mgpi in lower case.