MGP WRM2 Dokumentacja Strona 33

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 62
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 32
15-00044
Revision 5
April 2008
The publication, translation or reproduction, either party or wholly, of this document are not allowed without our written consent. 33
RMV Operating Instructions:
1) Start the RMV software by CLICKING on the CALIBRATION button on the RMV software connect the
DRM to the P.C. via the RS232 connection with a null modem connector, ensure dipswitches 1 and 2 on in
the ON position.
2) Set the correct Communication port
3) Select meter type, AMP+DRM
4) Verify that the communication between the meter and the PC is okay. If the communication is fail, change
the baud rate by clicking on the Setting, Communication port button, and select another baud rate. Then
click on Utility, Test to check again.
5) O
nce the communication is okay, use the Setting, Setup AMP button to configure the meter's working
parameters.
6) The Serial Number shown in the Setup Device screen refers to the ID number used in the communication
protocol and not the serial number of the instrument which is setup using the RMC (Rotem Meter
Calibration) or the AMP View software. All portable instruments telemetry ID will start with a ‘9’
followed by the unique identification number
M
eter Code for Telemetry (Area Monitors/Portable Instruments 9xxxxx)
Instrument Type Code
TelePole 0
Ram Ion 1
AMP-50 2
AMP-100 3
AMP-200 4
DRM-1 5
DRM-1D 6
Failed communication Communication OK
Przeglądanie stron 32
1 2 ... 28 29 30 31 32 33 34 35 36 37 38 ... 61 62

Komentarze do niniejszej Instrukcji

Brak uwag